WORK WITH THE BEST
Our new DT-3370 Series Heavy Duty Digital Clamp meters offer all the features you need to fit the way you work. It provides a full range of state of-the-art features to meet even the most demanding job requirements. The body and jaw perfectly fits in your and into tight areas of work. Meter controls are positioned so that current measurements can be done with one hand.
All the three clamp meters have improved based features such as a large, backlit display, true-RMS standards, CAT-IV safety ratings and a durably constructed housing. Have increased reading capability of 1000A and 1000V AC and DC.
These clamp meters confirm to CAT-III 1000V, CAT-IV 600V and are TUV & GS Approved. They confirm to EMC & LVD, EN:61326, EN:61010-1, EN:61010-02-031
Function | Range | Resolution | Accuracy |
DC Current | 400A | 0.1A | ±(3.0%±8Digits) |
1000A | 1A | ±(3.0%±8Digits) |
Function | Range | Resolution | Accuracy |
AC Current | 400A | 0.1A | ±(2.8%±8Digits) |
True RMS | |||
(50 to 60Hz) | 1000A | 1A | ±(3.0%±8Digits) |
Function | Range | Resolution | Accuracy |
DC Voltage | 400mV | 0.1mV | ±(0.8%±2Digits) |
4V | 0.001V | ±(1.5%±2Digits) | |
40V | 0.01V | ||
400V | 0.1V | ||
600V | 1V | ±(2.0%±2Digits) |
Function | Range | Resolution | Accuracy |
AC Voltage True RMS (50Hz to 400Hz) | 400mV | 0.1mV | ±(1.8%±20Digits) |
4V | 0.001V | ±(1.5%±5Digits) | |
40V | 0.01V | ||
400V | 0.1V | ||
600V | 1V | ±(2.0%±8Digits) |
Input Protection : 600V AC RMS.
Function | Range | Resolution | Accuracy |
Resistance | 400Ω | 0.1Ω | ±(1.0%±4Digits) |
4KΩ | 0.001KΩ | ±(1.5%±2Digits) | |
40KΩ | 0.01KΩ | ||
400KΩ | 0.1KΩ | ||
4MΩ | 1KΩ | ±(2.5%±5Digits) | |
40MΩ | 10KΩ | ±(3.5%±5Digits) |
Input Protection : 250V DC or 250V AC RMS
Function | Range | Resolution | Accuracy |
Capacitance | 4.000nF | 1pF | ±(5%±30Digits) |
40.00nF | 10pF | ±(5%±20Digits) | |
400.0nF | 0.1nF | ±(3%±5Digits) | |
4.000uF | 1nF | ||
40.00uF | 10nF | ||
400.0uF | 0.1uF | ±(4%±10Digits) | |
4.000mF | 0.001mF | ±(4.5%±10Digits) |
Input Protection : 250V DC or 250V AC RMS.
Function | Range | Accuracy (% of reading + digits) |
AC Voltage Frequency (Auto-ranging) | 10HZ-10KHZ | (1.0%±5Digits) |
Sensitivity : >15 VRMS
Function | Range | Resolution | Accuracy |
Frequency | 4Hz | 0.001Hz | ±(1%±5Digits) |
40Hz | 0.01Hz | ||
400Hz | 0.1Hz | ||
4kHz | 1Hz | ||
40kHz | 10Hz | ||
400kHz | 100Hz | ||
10MHz | 10kHz |
Sensitivity : > 5V RMS
Function | Range | Resolution | Accuracy |
Temperature (Type-k) (probe accuracy not included) | -20to | 1°C | ±(3%±5Digits) |
-4to1400°C | 1°F |
Input Protection : 250V DC or 250V AC RMS.
Function | Testing Condition | Reading |
Diode | Forward DCA is approx .1mA, open circuit Voltage | Forward voltage drop of Diode |
MAX.1.5V | ||
Continuity | Test current | Buzzer makes a long sound, While resistance is lessthan(35Ω) |
MAX.1.5mA |